What is Measurement Uncertainty
- Develop the specifications by writing a clear statement of exactly what is to be measured and the relationship between this and the parameters on which it depends. For example, if the measurement temperature has an influence on the result, the measurement temperature should also be defined.
- Develop a workflow diagram for the entire sampling, sample preparation, calibration, measurement, data evaluation and data transcription process (see Figure 1 for analytical sample testing).
- Identify and list sources of uncertainty for each part of the process or for each parameter. Possible sources for errors may be derived from non-representative sampling, operator bias, a wrongly calibrated instrument, lack of ideal measurement conditions, chemicals with impurities and errors in data evaluation.
- Estimate and document the size of each uncertainty, for example, as standard deviations or as RSDs. These data should be gathered from a series of measurements. Where experimental evaluation is impossible or impractical, the individual contributions should be estimated from whatever sources are available. Sources for this kind of estimation can be found in the supplier’s information or in the results of Interlaboratory studies or proficiency testing. The procedures and thoughts behind the way the contributions have been measured or estimated should be documented.
- Combine separate contributions in order to give an overall value. For example, where individual sources of uncertainty are independent, the overall uncertainty can be calculated as a multiple of the sum of squared contributing uncertainty components, all expressed as standard deviations. Computer software or spreadsheet programs can help to automate this calculation.
The whole procedure should be documented in such a way that sufficient information is available to allow the result to be reevaluated if new information or data become available. A complete documentation should include:
- A description of the methods used to calculate the measurement result and its uncertainty from the experimental measurements.
- The values and sources of all corrections.
- A list of all components of uncertainty with full documentation on how each of these was evaluated.
Measurement
What is a measurement?
What is not a measurement?
Uncertainty of measurement
What is uncertainty of measurement?
Expressing uncertainty of measurement
Error versus uncertainty
It is important not to confuse the terms ‘error’ and ‘uncertainty’.
- Error is the difference between the measured value and the ‘true value’ of the thing being measured.
- Uncertainty is a quantification of the doubt about the measurement result.
- Whenever possible we try to correct for any known errors: for example, by applying
- corrections from calibration certificates. But any error whose value we do not know is a source of uncertainty.
Why is uncertainty of measurement important?
You may be making the measurements as part of a:
- calibration - where the uncertainty of measurement must be reported on the certificate
- Test - where the uncertainty of measurement is needed to determine a pass or fail or to meet a
- Tolerance - where you need to know the uncertainty before you can decide whether the
- Tolerance is met ... or you may need to read and understand a calibration certificate or a written specification for a test or measurement.
Where do errors and uncertainties come from?
- The measuring instrument - instruments can suffer from errors including bias, changes due to ageing, wear, or other kinds of drift, poor readability, noise (for electrical instruments) and many other problems.
- The item being measured - which may not be stable. (Imagine trying to measure the size of an ice cube in a warm room.)
- The measurement process - the measurement itself may be difficult to make. For example measuring the weight of small but lively animals presents particular difficulties in getting the subjects to co-operate.
- 'Imported' uncertainties - calibration of your instrument has an uncertainty which is then built into the uncertainty of the measurements you make. (But remember that the uncertainty due to not calibrating would be much worse.)
- Operator skill - some measurements depend on the skill and judgement of the operator. One person may be better than another at the delicate work of setting up a measurement, or at reading fine detail by eye. The use of an instrument such as a stopwatch depends on the reaction time of the operator. (But gross mistakes are a different matter and are not to be accounted for as uncertainties.)
- Sampling issues - the measurements you make must be properly representative of the process you are trying to assess. If you want to know the temperature at the work-bench, don’t measure it with a thermometer placed on the wall near an air conditioning outlet. If you are choosing samples from a production line for measurement, don’t always take the first ten made on a Monday morning.
- The environment - temperature, air pressure, humidity and many other conditions can affect the measuring instrument or the item being measured. Where the size and effect of an error are known (e.g. from a calibration certificate) a correction can be applied to the measurement result. But, in general, uncertainties from each of these sources, and from other sources, would be individual ‘inputs’ contributing to the overall uncertainty in the measurement.
The general kinds of uncertainty in any measurement
- The measuring instrument - instruments can suffer from errors including bias, changes due to ageing, wear, or other kinds of drift, poor readability, noise (for electrical instruments) and many other problems.
Random or systematic The effects that give rise to uncertainty in measurement can be either :
- Random - where repeating the measurement gives a randomly different result. If so, the more measurements you make, and then average, the better estimate you generally can expect to get. or
- Systematic - where the same influence affects the result for each of the repeated measurements (but you may not be able to tell). In this case, you learn nothing extra just by repeating measurements. Other methods are needed to estimate uncertainties due to systematic effects, e.g. different measurements, or calculations.
The two ways to estimate uncertainties
- Type A evaluations - uncertainty estimates using statistics (usually from repeated readings)
- Type B evaluations - uncertainty estimates from any other information. This could be information from past experience of the measurements, from calibration certificates, manufacturer’s specifications, from calculations, from published information, and from common sense.
Eight main steps to evaluating uncertainty
The main steps to evaluating the overall uncertainty of a measurement are as follows.
- Decide what you need to find out from your measurements. Decide what actual measurements and calculations are needed to produce the final result.
- Carry out the measurements needed.
- Estimate the uncertainty of each input quantity that feeds into the final result. Express all uncertainties in similar terms.
- Decide whether the errors of the input quantities are independent of each other. If you think not, then some extra calculations or information are needed. (See correlation in Section 7.3.)
- Calculate the result of your measurement (including any known corrections for things such as calibration).
- Find the combined standard uncertainty from all the individual aspects. (See Section 7.2.)
- Express the uncertainty in terms of a coverage factor (see Section 7.4), together with a size of the uncertainty interval, and state a level of confidence.
- Write down the measurement result and the uncertainty, and state how you got both of these.
How to reduce uncertainty in measurement
- Calibrate measuring instruments (or have them calibrated for you) and use the calibration corrections which are given on the certificate.
- Make corrections to compensate for any (other) errors you know about.
- Make your measurements traceable to national standards – by using calibrations which can be traced to national standards via an unbroken chain of measurements. You can place particular confidence in measurement traceability if the measurements are quality-assured through a measurement accreditation
- Choose the best measuring instruments, and use calibration facilities with the smallest uncertainties.
- Check measurements by repeating them, or by getting someone else to repeat them from time to time, or use other kinds of checks. Checking by a different method may be best of all.
- Check calculations, and where numbers are copied from one place to another, check this too.
- Use an uncertainty budget to identify the worst uncertainties, and address these.
- Be aware that in a successive chain of calibrations, the uncertainty increases at every step of the chain.
Clients
Says
Thanks to the dedicated team for their expertise!

Technical Director
Calibration laboratory - USA
As a calibration laboratory, maintaining the highest standards of accuracy and reliability is critical. Lakshy Management Consultant Pvt. Ltd. were instrumental in helping us achieve ISO/IEC 17025 accreditation. Their team meticulously reviewed our calibration procedures, ensured our equipment met the necessary requirements, and helped us establish traceability to national standards. We were particularly impressed with their practical approach and their ability to explain complex technical concepts in a clear and understandable way. We highly recommend their services to any calibration laboratory seeking accreditation
Lakshy Management excels in ISO 17025 accreditation guidance.

CEO
Electrical Test Lab
“Achieving ISO/IEC 17025 accreditation was a major goal for our materials testing laboratory, but the process seemed daunting. Lakshy Management Consultant Pvt. Ltd. guided us every step of the way. Their consultants were incredibly knowledgeable about the standard and helped us develop a robust quality management system. They provided practical advice, helped us optimize our testing procedures, and ensured our documentation was impeccable. Thanks to their expertise, we sailed through the accreditation audit and are now confidently providing accurate and reliable testing services to our clients.”
Lakshy Management eased ISO 17025 implementation seamlessly and efficiently.

Lab Director
The largest Pharma Company
We were struggling to meet the stringent requirements of ISO/IEC 17025 for our medical testing laboratory. Lakshy Management Consultant Pvt. Ltd. provided the expertise and support we desperately needed. Their consultants helped us improve our risk management processes, implement effective internal audits, and ensure the competency of our laboratory personnel. They were always available to answer our questions and provide guidance, making the entire accreditation process much smoother than we anticipated. We are now a fully accredited laboratory and are confident in our ability to provide high-quality testing services to our patients
24 Hours Customer Care for ISO/IEC 17025 consulting / Accreditation: +91 9821780035
Contact Information
Our team of expert test & calibration experts will make ISO/IEC 17025 accreditation for your laboratory a reality. let’s get started!