Method Validation

Method validation is the process used to confirm that the analytical procedure employed for a specific test is suitable for its intended use. Results from method validation can be used to judge the quality, reliability and consistency of analytical results; it is an integral part of any good analytical practice. Analytical methods need to be validated or revalidated
The validity of a specific method should be demonstrated in laboratory experiments using samples or standards that are similar to unknown samples analyzed routinely. The preparation and execution should follow a validation protocol, preferably written in a step-by-step instruction format. Possible steps for a complete method validation are listed in Table 1. This proposed procedure assumes that the instrument has been selected and the method has been developed. It meets criteria such as ease of use; ability to be automated and to be controlled by computer systems; costs per analysis; sample throughput; turnaround time; and environmental, health and safety requirements.
ISO/IEC 17025 definition: ‘The confirmationby examination and the provision of objective evidencethat the particular requirements for a specific intended use are fulfilled’
What is validation?

Specific intended use

– analytical requirement
– what will test method be used for? why are measurements required?

Objective evidence

– experimental data from method validation study
– information on method performance

Confirmation

– comparison of requirement with experimental evidence
Can the method deliver results that are fit for a particular purpose?
Without validation there can be no assurance that results will be fit for purpose

During method development

– is the proposed method likely to be fit for purpose?

Before using any method for analysis of test samples

– including standard/published methods which have been validated by others
– verify suitability for analytical requirement
– verify own ability to match published data

Change of application/working environment/analyst

Following period of non-use

§5.4.1 General

– laboratory shall use appropriate methods and procedures for all tests within its scope

§5.4.2 Selection of methods

– methods must meet customer requirements
– preference for standard methods (international/national standards)
– ‘laboratory shall confirm that it can properly operate standard methods before introducing the tests’
– laboratory-developed methods may be used if they are validated

ISO definition

Types of method that require validation

– (standard -§5.4.2)
– non-standard
– laboratory designed/developed
– standard methods outside normal scope
– amplified/modified standard methods

Validation required to demonstrate that methods are fit for their intended use

– measurement uncertainty, detection limit, selectivity, linearity, repeatability/reproducibility, robustness

Why are the measurements required?

– what will the data be used for?

Analyte

– quantitative or qualitative
– class/total/available/species
– likely concentration

Sample

– nature (matrix)
– size

Other considerations

– destructive vs non-destructive test
– timescale
– cost

Selectivity/specificity

– am I measuring what I think I’m measuring? are there any interferences?

Precision (repeatability, reproducibility)

– how close are the results of replicate measurements made on the same sample?

Bias, recovery

– how close are the results to the ‘right’answer

Working range

– range of analyte concentrations that can be measured reliably
– limit of detection, limit of quantitation

Ruggedness/robustness

– control necessary for each stage of the procedure

DEPENDS ON THE CRITICALITY OF THE MEASUREMENT, SCOPE OF THE METHOD, LEVEL OF INFORMATION AND EXPERIENCE ALREADY AVAILABLE

Parameter Type of Analysis
Qualitative Major Component Trace Analysis Physical Property
Specficity / selectivity
Precision  
Bias  
Limit of Detection    
Limit of Quantitation      
Linearity/working range  
Ruggedness
Major component : analyte concentration in range 1-100%

Trace analysis : analyte concentration less than 100 mg/kg

Analyse data from method performance parameters

Are target values achieved?

– YES – method is fit for purpose
– NO – more development required

Method is validated by the declaration of fitness for purpose

Method validation is required to produce meaningful data

Both in-house and standard methods require validation/verification

Key requirement of ISO/IEC 17025

Validation should be a planned activity

– parameters required will vary with application

Validation is not complete without a statement of fitness for purpose

Clients
Says

24 Hours Customer Care for ISO/IEC 17025 consulting / Accreditation:  +91 9821780035

Projects Completed
0 +
On Time Completion
0 %
Employee's
0 +
Years Of Experience
0 +

Contact Information

Our team of expert test & calibration experts will make ISO/IEC 17025 accreditation for your laboratory a reality. let’s get started!

Register Now